Accredited calibration · Probes and clamps

Current probe, injection probe and EM clamp calibration

Transfer impedance, coupling and decoupling factor, insertion loss and probe attenuation measured on site under LAB Support Ltd.'s A2LA accredited ISO/IEC 17025 scope, by the LAB Support Southeast Asia team, with the certificate issued in your laboratory.

Why it matters

The probe converts what you inject or measure into a number

A current probe's transfer impedance turns a receiver reading into amperes. An injection probe or EM clamp's coupling factor sets the level of a bulk current injection or IEC 61000-4-6 test. If either has drifted, the whole test is offset. These devices are ferrite-based and sensitive to mechanical damage, so laboratories that move them between benches see the most drift. The measurement uses the calibration fixture the standard describes and a network analyser, which the team brings to your bench.

  • RF current probes and current sensors, transfer impedance and transfer admittance
  • BCI injection probes and calibration fixtures for ISO 11452-4 and MIL-STD-461 CS114
  • EM clamps and decoupling clamps for IEC 61000-4-6, impedance and coupling factor
  • Capacitive coupling clamps for IEC 61000-4-4, high-voltage and differential voltage probes
ParameterRangeCommon methods and standards
Current probe transfer impedance and admittance9 kHz to 1 GHzCISPR 16-1-2, calibration fixture, network analyser
Injection probe insertion loss in fixture9 kHz to 1 GHzISO 11452-4, MIL-STD-461 CS114, IEC 61000-4-6
EM clamp impedance, coupling and decoupling factor9 kHz to 1 GHzIEC 61000-4-6, network analyser
Decoupling clamp impedance and decoupling factor9 kHz to 1 GHzIEC 61000-4-6
Capacitive coupling clamp, EFT pulse at the clamp0.1 kV to 5 kVIEC 61000-4-4 with the generator, oscilloscope
Voltage probes, attenuation and bandwidthDC to 500 MHzOscilloscope and network analyser

The methods and standards shown are the ones most commonly requested. Within each parameter and range listed on the accredited scope, other calibration methods and standard editions can be supported under the flexible-scope provisions of LAB Support's A2LA accreditation. The measurement uncertainty for each parameter is stated on the certificate. Field calibration is included in the scope for every line above.

Instruments we calibrate

Probes and clamps we calibrate most often

From our own calibration records. Any probe or clamp within the scope can be calibrated; these are the ones we see most.

Teseq KEMZ 801 / 801A EM clampLuthi EM101 EM clampMEB KEMZ 801Teseq MD 200, MD 300 targets and probesTeseq CSP 9160A current probeTeseq CIP 9136A injection probe Fischer F-33-2, F-51, F-61 current probesFischer F-120, F-140 injection probesSchwarzbeck TK 9420 voltage probeSchwarzbeck MDS 21 decoupling clampTeseq CCI / EM TEST CCI capacitive clampHaefely IP4A injection probeTeseq high-voltage differential probesand others
On site

How a probe and clamp calibration runs

Send the listModels, serial numbers and the fixture you own, if any. We bring fixtures for the common probe families.
We come to youThe team brings a calibrated network analyser, calibration kit, probe fixtures and terminations to your laboratory.
Measure in placeEach probe is measured in its fixture across the standard's frequency points. Capacitive clamps are checked with your EFT generator.
Certificate on siteISO/IEC 17025 certificate with tabulated transfer impedance or coupling factor, plots, uncertainty and traceability.
Frequently asked

Probe and clamp calibration questions

Do you need my calibration fixture?

If you have the manufacturer's fixture we use it, because the standard defines the result in that fixture. For the common Teseq, Fischer and Schwarzbeck probes the team carries its own fixtures.

Can BCI probes for automotive and military tests be calibrated?

Yes. Injection probes and their fixtures for ISO 11452-4 and MIL-STD-461 CS114 are measured for insertion loss in the fixture, which is what the test procedures use to set the forward power.

How often should EM clamps be calibrated?

Most laboratories use 12 to 24 months. Clamps with damaged ferrites, cracked housings or replaced connectors should be recalibrated immediately, since the coupling factor changes with the ferrite path.

Are capacitive coupling clamps calibrated alone or with the generator?

With the generator, as IEC 61000-4-4 describes: the burst waveform is verified at the clamp with a transducer plate and oscilloscope. The generator itself is calibrated at the same visit if required.

Tell us what you need to test

We reply within one working day with a quote or a plan.

Contact us