Voltage division factor, isolation and impedance of coupling/decoupling networks measured in your laboratory under LAB Support Ltd.'s A2LA accredited ISO/IEC 17025 scope, with a certificate issued on site by the LAB Support Southeast Asia team.
In an IEC 61000-4-6 conducted immunity test the CDN defines the common-mode impedance and the voltage that reaches the equipment under test. If its division factor has drifted, every test level in the laboratory is wrong by the same amount, whichever generator or amplifier drives it. CISPR 16-1-2 and IEC 61000-4-6 both expect the network to be characterised on a schedule, and assessors ask for the certificate.
| Parameter | Range | Common methods and standards |
|---|---|---|
| Voltage division factor | 9 kHz to 1 GHz | IEC 61000-4-6, network analyser with calibration kit |
| Isolation (decoupling) | 9 kHz to 1 GHz | IEC 61000-4-6, network analyser |
| EUT port impedance | 9 kHz to 1 GHz | IEC 61000-4-6 Annex A, 150 Ω adapters |
| 150 Ω to 50 Ω adapters | 9 kHz to 1 GHz | Insertion loss and return loss |
| EM clamps and decoupling clamps | 9 kHz to 1 GHz | Impedance, coupling and decoupling factor |
| Burst and surge CDNs | With the generator | IEC 61000-4-4 and -4-5 pulse parameters at the CDN output |
The methods and standards shown are the ones most commonly requested. Within each parameter and range listed on the accredited scope, other calibration methods and standard editions can be supported under the flexible-scope provisions of LAB Support's A2LA accreditation. The measurement uncertainty for each parameter is stated on the certificate. Field calibration is included in the scope for every line above.
Taken from our own calibration records. Any make of coupling network within the scope can be calibrated; these are simply the ones we see most.
Most accredited laboratories use a 12 to 24 month interval, set by their own quality system and the assessor's expectations. Networks that are moved between benches, have replaced connectors or show test-level anomalies should be checked sooner.
Yes. For burst, surge and ring-wave CDNs the pulse parameters are verified at the CDN output with the generator that drives it, as IEC 61000-4-4 and IEC 61000-4-5 describe. The generator itself is calibrated in the same visit if you wish.
Yes. Field calibration is written into LAB Support Ltd.'s A2LA scope for these parameters, so a certificate issued in your laboratory carries the same accreditation as one issued in Japan.